Assessment of the Quality of Color Reproduction of 3D Medical Images
Abstract:
This article discusses the problem of assessing the quality of color reproduction in 3D medical images. In the context of medical images, accurate and appropriate color reproduction plays an important role in correct interpretation and diagnosis. The article provides an overview of existing standards and metrics used to evaluate color reproduction in the field of medical imaging. Aspects such as the choice of color space, the accuracy of color transmission, and consistency between different display devices, human perception of color and other factors affecting the quality of color reproduction are considered. Testing and improvement of the qualitative characteristics of the color reproduction of the monitor using software and hardware to standardize its color parameters has been performed. The color distance of the monitor in the CIE L*a*b* color space before calibration was 14.84, and after calibration 1.23.
Keywords:
Assessment, color reproduction, medical image, standards, calibration of monitors, endoscopic surgery